EasyManua.ls Logo

Optek CONTROL 200 - Page 98

Optek CONTROL 200
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH D-45356 Essen Germany
page 76
Span Value
mS/cm and Adjust are shown alternately on the upper display line. Use the UP
and
DOWN keys to adjust the displayed reading to the required span value – see
Table above.
Zero value.
Note. If A: Analyzer Type is set to Conc, the displayed units are those set in the
A: Concentration or A: User Units parameters – see Section 5.2.3.
Zero Value
Span value.
mS/cm and Adjust are shown alternately on the center display line. Use the and
keys to adjust the displayed reading to the required zero value – see Table
above.
Note. If A: Analyzer Type is set to Conc, the displayed units are those set in the
A: Concentration or A: User Units parameters – see Section 5.2.3.
Default Output
Select the system reaction to failure:
Off – Ignore failure and continue operation.
On – Stop on failure. This drives the analog output to the level set in the
Default Val frame below.
Hold – Hold the analog output at the value prior to the failure.
Default Value
The level to which the analog output is driven if a failure occurs.
Set the value between 0.00 and 22.00mA.
Output 2 configuration (and Outputs 3 and 4 if option board fitted
and analog features enabled) is identical to Output 1 configuration.
Option board fitted and analog features enabled – see Section 5.2.6.
Option board fitted and Serial Communications feature enabled –
see Suppl. Manual PROFIBUS Datalink Description (IM/PROBUS).
Single input analyzer and option board not fitted – see Section 5.2.7.
Dual input analyzer and option board not fitted – see Section 5.2.8.
AO1: Zero Value
100.0
uS/cm
0.0
mS/cm
AO1: Default
-----
mA
Hold
On
Off
AO1: Default Val
22.00
mA
AO1: Span Value
100.0
mS/cm
0.0
uS/cm
Config. Output 1
Config. Output 2
CONFIG. SERIAL
CONFIG. CLOCK
CONFIG. SECURITY
CONFIG. CONTROL
Config. Output 1
On
Off or Hold

Table of Contents

Related product manuals