115
Measurement review
On the main quenching measuring screen, the lower right hand side window logs all of
the measurements made, in the current measuring session, for the current file name. One
can scroll through measurements by touching the screen. If one touches the top of the
review area window, the earliest measurements may be reviewed. If one touches the
bottom of the review area window, then the latest measurement can be reviewed. The
session remains current as long as the file name remains the same, and one does not leave
the quenching protocol. It will stay intact if one goes to different screens within the
quenching protocol. If one turns off the instrument, changes the file name, or goes to a
different protocol, then the session is ended.
After a session is ended, data, and traces may still be viewed by going to the main menu,
selecting “Data Management”, highlighting the file of interest with your finger, and
pressing view. If a trace is saved with the quenching data, the trace may be seen by using
the “Line up” and “Line down” arrow keys to go to the last measurement of a quenching
trace. It there is a trace saved, a low resolution trace will appear in the lower left hand
corner of the screen. For more information, go to the “Data Management” section of the
manual.
Measuring data and traces are also retained with the current file name in the measuring
file, and it may be viewed with Excel on a PC computer. It does not disappear like the
measurement review session data in the window.