Loading and saving preset measuring routines
When the parameters have been adjusted for a specific set of conditions, the test
parameter setup can be saved by pressing “Save preset” while on the Test Setup Screen.
A key board screen will appear, and the test set up can be named, and it is saved, when
“Ent” is pressed. In Fact, changes that are made to the “Test Collection Control” screen,
the “Fv/Fm Measurement Pulse Setup” screen, and the “Fv/Fm Test Setup” screen can all
be saved in a single measuring file by pressing “Save Preset” on the Fv/Fm Test Set up
screen. The changes from other screens may be done in any order, and saved from this
screen when complete. If parameters are changed, and a new file name is not given to this
new set of parameters, then the changes will be made to the existing preset file. Follow
the directions listed above to save measuring parameters in files.
To load previously created tests, press “Load preset”. Press the test of interest. After it is
highlighted in light blue, press load. To erase a test, highlight the test and press “Erase”.
The number of tests that can be saved is almost unlimited. Use the up and down arrows to
scroll through the tests if there are more that are not shown on the screen. Press “Exit” if
there is no action.
Fv/Fm Test Setup Screen 1p Configuration Loader Screens
Loading measuring parameter preset files –
Highlight test by touch and press load.
Arrows appear when there are more tests than can
be viewed at one time on the screen.
Copy only works when an SD card is inserted
into the side of the OS1p.