MX40 Release C.01 Compatibility
2-2 Installation
MX40 Release C.01 Compatibility
IntelliVue Information Center
The MX40 (865350/865351) is compatible for use with IntelliVue Information Center
Release N/N.01 only. Release L and M are not supported.
The MX40 (865352) is not compatible with IntelliVue Information Center Release N
or earlier.
Patient Information Center iX
The MX40 (865350/865351/865352) with ECG only or optional FAST SpO2 remains
compatible with IntelliVue Information Center iX Release A.02.08 or latest release.
The MX40 (865350/865351/865352) with optional Masimo SET technology is
compatible for use with IntelliVue Information Center iX Release B.02.15 or latest
release.
Note— 865350 and 865351 Smart-hopping devices operate on Smart-hopping
Network infrastructure only. 865352 WLAN devices operate on 802.11 wireless
network infrastructure only.
MX40 Short-range Radio
The MX40 (865350/865351) is compatible for use with IntelliVue Patient Monitors
Release G or latest release when wirelessly connected via Short-range radio.
The MX40 (865352) is compatible for use with IntelliVue Patient Monitors Release J
or latest release when wirelessly connected via Short-range radio.
The MX40 is compatible for use with IntelliVue Cableless Measurements Release
C.0 or latest release.
MX40 and Smart-hopping Infrastructure
The MX40 (865350/865351) is compatible for use with Access Point Controller
862147, Release B.00.19 or latest release and Access Point Controller 865346, Release
C.00.04 or latest release.
MX40 Patient Cable
The MX40 Patient Cables are compatible for use with the X2, X3, MMS, and MMX
IntelliVue Patient Monitor platforms.