Battery Specifications
10-18 Safety Standards & Specifications
Note The battery capacity of re-chargeable batteries degrades over time
and number of recharge cycles. Toward the end of its useful life, the battery
capacity may be reduced by 30%. If this reduced battery life is unacceptable
based on your use model, Philips recommends replacing the rechargeable
battery sooner.
Nominal Current
Operating Mode
Nominal Current
(p/n 865350)
Nominal Current
(p/n 865351)
Nominal Current
(p/n 865352)
ECG Only FAST (Display
inactive)
51.5 mA @ 3.6V 51.0 mA @ 3.6V 51.5 mA @ 4.1V
ECG/SpO
2
FAST
Continuous
(Display inactive)
120 mA @ 3.6V 116 mA @ 3.6V 120 mA @ 4.1V
ECG Only Masimo (Display
inactive)
51.5 mA @ 3.6V 51.0 mA @ 3.6V 51.5 mA @ 4.1V
ECG/SpO
2
Masimo
Continuous
(Display inactive)
122 mA @ 3.6V 118 mA @ 3.6V 123 mA @ 4.1V
Lithium-ion Battery Charge Time
Definition Charging Method
Charge
Time
Battery pack charge
time from 90%
depletion state
The Lithium-ion Battery Pack is charged on
a separate external charging station. It
must be removed from the MX40 to
charge.
6.5 hours