EasyManua.ls Logo

Philips IntelliVue MX40 - Electrosurgery Interference;Defibrillation;Electrostatic Discharge; Restart Time

Philips IntelliVue MX40
184 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electromagnetic Compatibility
10-14 Safety Standards & Specifications
Frequency of
Transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Equation d = 1.2√P d = 1.2√P d = 2.3√P
Rated max. output
power of transmitter
Separation distance Separation distance Separation distance
0.01 W 0.1 m 0.1 m 0.2 m
0.1 W 0.4 m 0.4 m 0.7 m
1 W 1.2 m 1.3 m 2.3 m
10 W 3.8 m 3.8 m 7.3 m
100 W 12.0 m 12.0 m 23.0 m
Electrosurgery Interference/Defibrillation/Electrostatic
Discharge
The equipment returns to the previous operating mode within 10 seconds
without loss of any stored data. Measurement accuracy may be temporarily
decreased while performing defibrillation. This does not affect patient or
equipment safety. Do not expose the equipment to x-ray or strong magnetic
fields (MRI). The MX40 is not for use during electrosurgery.
Restart Time
After power interruption, an ECG wave will be shown on the display after 30
seconds maximum.

Table of Contents

Other manuals for Philips IntelliVue MX40

Related product manuals