MX40 Test & Inspection Matrix
Test and Inspection 3-3
Test Block
Name
Test or "Inspection" to Perform
What to Record
on Service
Record
9. If the MX40 has the Short-Range Radio option, establish
communication between the MX40 and either the patient
monitor or a cableless measurement device, depending on the
chosen use model. . If assigned to a patient monitor, an ECG
waveform should be visible on the monitor. The display on the
MX40 will be:
10. If assigned to a cableless measurement device, initiate a
measurement and view it at the Information Center.
Expected Test Results: Expected answer to all is "yes". If so,
Performance test is passed.
Revision
Check:
Check the revision of the software/firmware in the Device Info.
screen. Check the INOP Area for either an "SpO
2
Equip Malf"
or "Resp Equip Malf" message which indicates an SpO
2
or
Resp upgrade failure. The revision reported should match the
revision loaded. You may also check the Status Log at the
Information Center.
Note: If the device has SpO2 hardware installed (option S02
or S04), after an upgrade if SpO2 is in Manual or Auto
measurement mode, the SpO2 revision information will display
as ??.??.??. To view the SpO2 revision information go to the
SmartKeys and select “Start SpO2”, then return to the Device
Info screen.
Expected Test Results: Expected answer is "yes". If so,
Revision Check test is passed.
RC:P or
RC:F
where P=Pass
F=Fail