Configuration Mode
Operating Modes 4-31
Setting Description MX40 with IIC N
MX40 with IIC
L/M
MX40 with IIC iX
Alarm Sounds: Sets MX40 alarm sound
type to Traditional
(Carenet) or ISO.
Traditional, ISO Traditional, ISO Traditional, ISO
Alarms On: Enable: All MX40/IIC
Release N features
available.
Disable: MX40 operates
as if connected to IIC
Release L/M.
Disable, Enable
Note — This
configuration item is
not available when
operating with IIC iX.
Disable Enable (cannot
disable - MX40 is the
alarm manager)
Lock Time Sets the default
automatic lock time
(minutes).
1, 2, 5, 15, 30 1, 2, 5, 15, 30 1, 2, 5, 15, 30
Unit Defaults:
The table below lists the settings that are configured using the SmartKeys
menu:
Setting Description MX40 with IIC N
MX40 with IIC
L/M
MX40 with IIC iX
Alarm Volume
for Off Network:
Sets the default alarm
volume when the device
goes off network
10 only 10 only 10 only
Inop Reminder: Inop reminders on or off Set at IIC On, Off Set at IIC iX
Inop Severity:
• ECG Leads
Off
• Replace
Battery
• Leadset
Unplugged
• Invalid
Leadset
Sets the severity of the
"ECG Leads Off",
"Replace Battery" and
“Leadset Unplugged”
INOP conditions
Set at IIC Red, Yellow, Cyan Set at IIC iX
• Screen On
Time
Sets default screen on
time (minutes)
1, 2, 5, 15, 30 1, 2, 5, 15, 30 Set at IIC iX
• Trend Group
Sets default trend group
type: Standard (no
ST/QT) or Cardiac
(includes ST/QT)
Standard (Cardiac not
available)
Standard (Cardiac not
available)
Standard, Cardiac
• Trend Interval
Sets default trend
interval time
1, 5, 10, 15, 30, 1
hour, 2 hours
1, 5, 10, 15, 30, 1
hour, 2 hours
1, 5, 10, 15, 30, 1
hour, 2 hours