Part 2 Isolating and Solving Instrument Problems
196
Troubleshootin
the Instrument
“Local Recorder
door open” sta-
tus message
Recorder door is open.
The Recorder is availa-
ble, but is not ready for
recording.
Close the door.
“Local Recorder
out of paper”
status message
The Recorder is out of
paper.
The Recorder is availa-
ble, but is not ready for
recording.
Load paper into the
Recorder (refer to the
Operator’s Manual for the
XE-50p Chart Recorder
for details.
“No Local
Recording
(Printhead Tem-
perature)” sta-
tus message
The printhead is too warm
(> 68°C).
Switch off the Recorder
and allow it to cool down
for a few minutes, then try
again. If the problem per-
sists, replace the Recorder.
The printhead is too
cold(< 0°C).
(For instance, the
Recorder may have been
tranferred from storage,
where the temperature
was below 0°C).
The ambient temperature
in the room where the
Recorder is used should
be above 0°C. Allow the
Recorder to warm up for a
few minutes, then try
again.
If the problem persists,
replace the Recorder.
“Local Recorder
disconnected”
status message
The Recorder is discon-
nected from the Monitor
or switched off
• Ensure the Recorder
cable is securely
connected to the sockets
on the Monitor and on the
Recorder.
• Ensure that the Recorder
is switched on.
“Local Recorder
connected”
The Recorder is connected
and available.
“No local alarm
recording avail-
able” status
message
Local alarm recording has
been selected, but the
Recorder is not available.
Ensure that the Recorder
is switched on and con-
nected to the Monitor.