Part 2 Isolating and Solving Instrument Problems
196
Troubleshootin
g
the Instrument
“Local Recorder
door open” sta-
tus message
Recorder door is open.
The Recorder is availa-
ble, but is not ready for
recording.
Close the door.
“Local Recorder
out of paper”
status message
The Recorder is out of
paper.
The Recorder is availa-
ble, but is not ready for
recording.
Load paper into the
Recorder (refer to the
Operators Manual for the
XE-50p Chart Recorder
for details.
“No Local
Recording
(Printhead Tem-
perature)” sta-
tus message
The printhead is too warm
(> 68°C).
Switch off the Recorder
and allow it to cool down
for a few minutes, then try
again. If the problem per-
sists, replace the Recorder.
The printhead is too
cold(< 0°C).
(For instance, the
Recorder may have been
tranferred from storage,
where the temperature
was below 0°C).
The ambient temperature
in the room where the
Recorder is used should
be above 0°C. Allow the
Recorder to warm up for a
few minutes, then try
again.
If the problem persists,
replace the Recorder.
“Local Recorder
disconnected”
status message
The Recorder is discon-
nected from the Monitor
or switched off
Ensure the Recorder
cable is securely
connected to the sockets
on the Monitor and on the
Recorder.
Ensure that the Recorder
is switched on.
“Local Recorder
connected”
The Recorder is connected
and available.
“No local alarm
recording avail-
able” status
message
Local alarm recording has
been selected, but the
Recorder is not available.
Ensure that the Recorder
is switched on and con-
nected to the Monitor.