EasyManua.ls Logo

Philips M3015A - M3015 A

Philips M3015A
293 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preventive Maintenance Tasks
124
Maintainin
g
the Instrument
M3015A
NOTE 1. Allow 5 seconds between individual service procedures in order to ensure stable
equipment conditions.
2. When certain Monitor procedures are running (for example, AutoZero or purging),
service procedures are not possible and trying to start them will result in a message
“Service Operation Failed” in the status line on the Monitor. Wait until the Monitor has
completed the current operation, then restart the service procedure.
Preventive maintenance comprises the following activities:
Perform visual inspection (see “Inspecting the Instrument” on page 122)
Check the operating time for the pump and CO
2
scrubbers (Pump OpTime).
If required, (operating time >15,000 hours), replace the pump and the CO
2
scrubber (units
with serial number prefix DE020xxxxx only)
, and reset the operating time counter.
Perform barometric pressure check and calibrate, if necessary, see “Barometric Pressure
Check and Calibration” on page 149.
Perform leakage check, see “Leakage Check” on page 150.
Perform Pump Check
Perform flow check and calibrate, if necessary, see “Flow Rate Check and Calibration” on
page 151.
Perform Noise Check
Perform CO
2
calibration check; check the date of the last calibration and calibrate, if neces-
sary, see “CO
2
Gas Measurement Calibration Check” on page 152.
If calibration done, perform CO
2
calibration verification using 2nd calibration gas
If M3015A has been opened, perform a system safety test, see “Patient Safety Checks” on
page 159.
Tools Required
Standard tools, e.g. screwdriver, tweezers
ESD protection equipment
Electronic flowmeter, M1026-60144
Gas calibration equipment
Cal 1 gas 15210-64010 (5% CO
2
)
Cal 2 gas 15210-64020 (10% CO
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
Required Replacement Part
Pump assembly and CO
2
scrubber (M3015-29303)

Table of Contents