Troubleshooting the Installed Instrument
Troubleshootin
the Instrument
211
SpO
2
EQUIP MALF Measurement defective. Exchange the
Measurement Server.
The Measurement Server should be investigated
technically as soon as possible.
TEMP EQUIP MALF Measurement defective. Exchange the
Measurement Server or Measurement Server
Extension.
The Measurement Server or Measurement Server
Extension should be investigated technically as
soon as possible.
CO
2
EQUIP MALF Measurement defective. Exchange the
Measurement Server Extension or (for mainstream
CO
2
) the transducer.
BAD SERVER LINK You cannot use this combination of Monitor,
Measurement Server and cable.
Switch off the monitor and contact your biomedical
department.
BAD SERVER LINK
plus
“Measurement Server
Revision not supported”
status message in red.
An M3000A Measurement Server with revision A
software is connected to an M3046A Monitor with
a software revision of B or later. This combination
does not allow monitoring.
“Some measurements are
not supported by the
Monitor” prompt message
A measurement extension (M3015A or M3016A) is
connected to an M3046A M3 monitor.
No measurements (for example CO
2
, 2nd Press/
Temp) from the Measurement Extension are
available.
If the extra measurements are required, consider
using a monitor that supports these measurements.
Message What To Do