882 Video Test Instrument User Guide (Rev A.35) 475
PDAX:RPTG
Enable or disable RePoRT Generation for the Pixel
Data Analyzer test
Class Signal analyzer - Pixel data
Description Enable or disable the generation of an HTML report for the results of the pixel data
analysis test.
Command syntax PDAX:RPTG boolean
boolean
0 (disable), 1 (enable) report generation
Query response PDAX:RPTG?
Query response boolean
Example
PDAX:FRMS 100 //sets the number of frames to analyze in the pixel
data analysis test to 100
PDAX:MXER 1000 //sets the maximum number of errors to record in pixel
data analysis test to 1000
PDAX:RPTG 1 //enables the generation of reports for the ppixel
data analysis test
PDAX:REFG 0 //enables auto reference capturing for the pixel
data analysis test
PDAX:CAPF //captures a frame of pixel data for the pixel
data analysis test
PDAU: //runs the pixel data analysis test
Related commands PDAX:MXER, PDAX:FRMS, PDAX:REFG, PDAX:CAPF, PDAU