882 Video Test Instrument User Guide (Rev A.35) 251
Deleting a test sequence
You can delete a test sequence using the command line interface.
To delete a test sequence using the command line interface:
1. Establish a session with the generator using either HyperTerminal over a serial
connection or Telnet over an Ethernet LAN. See “Establishing a terminal session with
the 882” on page 30 or “Establishing a Telnet session with the 882” on page 33.
2. Set the sequence path to the medium and location you want to query. For example, if
you want to delete the sequence files in the generator’s flash memory, enter the
following command:
SEQP /tffs0/Library/Sequence
To delete sequence files on the PC card, enter the following command:
SEQP /card0/Library/Sequence
3. Delete the sequence by entering the following commands:
SEQK seqname
ALLU