250 Chapter 8 Working with Test Sequences
Before:
...
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0659.xml</FMT>
<IMG>SMPTEBar.img</IMG>
<DELY>+3.0000000E+00</DELY>
</STEP>
...
After (modified text in bold):
...
<STEP>
<FMT>/tffs0/Library/TestAnalog/DMT0856.xml</FMT>
<IMG>PulseBar.img</IMG>
<DELY>+4.0000000E+00</DELY>
</STEP>
...
4. Save the text file as an XML file.
5. Using the Generator FTP Browser, download the format XML file from your PC to the
generator. See ā€œCopying files from a PC to a 882ā€ on page 23 for instructions.