14.2. Autozero
The software features an autozero function (Auto) ensuring precise mass
indication. This function automatically controls and corrects zero indication.
There are, however, some cases when this function can be a disturbing factor
for the measuring process; e.g. very slow placing of a load on the weighing pan
(load adding). In such case, it is recommended to disable the function.
Procedure:
Enter <P1.rEAd> submenu.
AUTO no - autozero disabled
AUTO YES - autozero enabled
Return to the weighing mode saving introduced modifications.
14.3. Tare
Function enables setting appropriate parameters related with taring.