- 54 -
21.4.16. Send all implemented commands
Format: PC CR LF
Response: PC_- >_Z,T,OT,UT,S,SI,SU,SUI,C1,C0,CU1,CU0,K1,K0,NB,PC -
command carried out, all implemented commands have been sent.
21.5. Manual Printout / Automatic Printout
It is possible to generate printouts either manually or automatically.
• Manual printout is generated for stable weighing result. Load the platform,
wait for a stable result and press
key.
• Automatic printout is generated for stable weighing result. Load the
platform, wait for a stable result. No key needs to be pressed.
Caution:
For verified balance option of temporary weighing results printout is disabled.
Format:
1 2 3 4 -12 13 14 15 16 17 18
stability
marker
space character mass space unit CR LF
Stability marker
[space] if measurement result stable
[?] if measurement result unstable
[^] if high limit is out of range
[v] if low limit is out of range
character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point, right justification
unit
3 characters, left justification
Command
3 characters, left justification
Example 1:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF - printout generated upon pressing
ENTER/PRINT key.
Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - printout generated upon pressing
ENTER/PRINT key.