- 48 -
K0
Unlock balance keypad
NB
Give balance serial number
PC
Send all implemented commands
Caution:
1. Each command must end with CR LF characters;
2. Wait before sending another command until the former answer has been
received, otherwise the answers may be lost.
21.3. Response Format
On receipt of a command, the indicator responds as follows:
XX_A CR LF
command understood and in progress
XX_D CR LF
command carried out (appears only after the XX_A command)
XX_I CR LF
command understood but not accessible at this moment
XX _ ^ CR LF
command understood but max threshold is exceeded
XX _ v CR LF
command understood but min threshold is exceeded
ES_CR LF
command not recognised
XX _ E CR LF
time limit exceeded while waiting for stable measurement result
(time limit is a characteristic balance parameter)
XX - name of a sent command
_ - space
21.4. Commands Overview
21.4.1. Zeroing
Format: Z CR LF
Response options:
Z_A CR LF
Z_D CR LF
command understood and in progress
command carried out
Z_A CR LF
Z_^ CR LF
command understood and in progress
command understood but zeroing range is exceeded
Z_A CR LF
Z_E CR LF
command understood and in progress
time limit exceeded while waiting for stable measurement result
Z_I CR LF
command understood but not accessible at this moment

Table of Contents