23. TROUBLESHOOTING
Problem Cause Solution
Battery (batteries) discharged,
Connect it to the mains, charge
the battery (batteries)
The balance does
not switch on
No batteries (batteries not
installed or installed incorrectly)
Check if batteries are installed
correctly (polarization)
The balance
switches off
automatically
't1' parameter set to 'YES' value
(the balance switches off
automatically)
In 'Misc' menu change <5.4.t1>
parameter setting to 'no' value.
During switching
on, message 'LH'
is displayed
Weighing pan loaded during
switching on
Unload the weighing pan. Zero
indication is displayed.
24. ERROR MESSAGES
Err2
- Value beyond zero range.
Err3
- Value beyond tare range.
Err4
-
Adjustment weight or start mass out of range (±1% for
adjustment weight, ±10 for start mass).
Err5
- Single part mass value is lower than reading unit
value.
Err8
- Tarring / zeroing / start mass determination /
adjustment operation time exceeded.
null
- Zero value from converter.
FULL2
- Weighing range exceeded.
LH
- Start mass error, indication out of range
(-5% – +15% of start mass).
5–FULL
- Limited screen capacity, display of full indication value
for totalizing function is impossible.
Caution:
1. Errors: You are informed about display of Err2, Err3, Err4, Err5, Err8,
null errors by short sound signal (1 second).
2. You are informed about display of FULL2 error by continuous sound
signal lasting as long as excess load remains on the weighing pan.