- 44 -
Adjustment process can be aborted at any time. To do that, press
key. The following message is displayed:
Caution:
1. It is crucial to carry out balance adjustment with unloaded weighing pan!
2. If adjustment process takes more than 15 seconds then <Err8> error
message is displayed and short sound signal is emitted. Press
key
and carry out adjustment procedure again. Remember to maintain stable
ambient conditions!
Return to the weighing mode saving introduced modifications.
18.2. Start Mass Determination
If the balance does not require adjustment or you do not have suitable amount
of weights, you can determine balance start mass.
Procedure:
Enter <P6.CAL> submenu. Next: