21.4.9. Switch on continuous transmission in basic measuring unit
Format: C1 CR LF
Response options:
C1_I CR LF
command understood but not accessible at this moment
C1_A CR LF
MASS FRAME
command understood and in progress
response: mass value in basic measuring unit
Response format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
21.4.10. Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Response options:
C0_I CR LF
command understood but not accessible at this moment
C0_A CR LF
command understood and carried out
21.4.11. Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Response options:
CU1_I CR LF
command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
command understood and in progress
response: mass value in current measuring unit
Response format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
stability
marker
space character mass space unit CR LF