- 55 -
Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - printout generated upon pressing
ENTER/PRINT key.
21.6. Continuous Transmission
For continuous transmission the balance provides option of mass
measurement printout in basic unit and in additional unit. The mode can be
activated with command sent via interface, or by setting respective parameter
values.
Format is valid for <P2.Prnt> parameter set to CntA value:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
Stability marker
[space] if measurement result stable
[?] if measurement result unstable
[^] if high limit is out of range
[v] if low limit is out of range
character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point, right justification
unit
3 characters, left justification
Command
3 characters, left justification
Format valid for <P2.Prnt> parameter set to Cntb value:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
character
marker
space character Mass space unit CR LF
Stability marker
[space] if measurement result stable
[?] if measurement result unstable
[^] if high limit is out of range
[v] if low limit is out of range
character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point, right justification
unit
3 characters, left justification
Command
3 characters, left justification