- 53 -
21.4.12. Switch off continuous transmission in current measuring unit
Format: CU0 CR LF
Response options:
CU0_I CR LF
command understood but not accessible at this moment
CU0_A CR LF
command understood and carried out
21.4.13. Lock balance keypad
Format: K1 CR LF
Response options:
K1_I CR LF
command understood but not accessible at this moment
K1_OK CR LF
command carried out
Caution:
Command is not saved on balance restart.
21.4.14. Unlock balance keypad
Format: K0 CR LF
Response: K0_OK CR LF - command carried out
21.4.15. Give balance serial number
Format: NB CR LF
Response options:
NB_A_”Nr fabryczny” CR LF
command understood, response: serial number
NB_I CR LF
command understood but not accessible at this moment
”nr fabryczny – serial number of the device. Inserted in between inverted
commas.
Example:
NB CR LF – command sent from a computer
NB_A_”123456” CR LF – balance serial number - 123456