- 49 -
 
21.4.2.  Tarring 
 
Format: T CR LF 
 
Response options: 
 
T_A CR LF 
T_D CR LF 
command understood and in progress
 
command carried out 
T_A CR LF 
T_v CR LF 
command understood and in progress
 
command understood but taring range is exceeded 
T_A CR LF 
T_E CR LF 
command understood and in progress
 
time limit exceeded while waiting for stable measurement result 
T_I CR LF 
command understood but not accessible at this moment 
 
 
21.4.3. Give Tare Value 
 
Format:  OT CR LF 
 
Response:  OT_TARA CR LF - command carried out 
 
Response format: 
 
1 2  3  4  5-6  7-15  16  17 18 19  20  21 
T T  space 
stability 
marker 
space tare space  unit  CR LF 
 
Tare  - 9 characters, right justification 
Unit  - 3 characters, left justification 
 
 
21.4.4. Set tare 
 
Format:  UT_TARA CR LF, where TARA - tare value 
 
Response options: 
 
UT_OK CR LF 
command carried out 
UT_I CR LF 
command understood but not accessible at this moment 
ES CR LF 
command not recognised (tare format incorrect)
 
 
Caution: 
Use dot in tare format as decimal point.