21.4.5. Send stable measurement result in basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
S_E CR LF
command understood and in progress
time limit exceeded while waiting for stable measurement result
S_I CR LF
command understood but not accessible at this moment
S_A CR LF
MASS FRAME
command understood and in progress
response: mass value in basic measuring unit
Response format:
1 2-3 4 5 6 7-15 16 17 18 19 20 21
S space
stability
marker
space character mass space unit CR LF
Example:
S CR LF – command sent from a computer
S _ A CR LF – command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out,
response: mass value in basic measuring unit.
21.4.6. Immediately send measurement result in basic measuring unit
Format: SI CR LF
Response options:
SI_I CR LF
command understood but not accessible at this moment
MASS FRAME
immediate response: mass value in basic measuring unit
Response format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
Example:
S I CR LF – command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,
immediate response: mass value in basic measuring unit