Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-15
Cedex Bio analyzer 5 Daily operation
Preparing the system
4
Prepare the reagent disk. 1. Choose .
2. Open the main cover.
3. Remove the disk from its container.
4. Place the disk on the analyzer. Make sure the
arrow indicators on the disk cover are in the
front.
5. Close the main cover.
After the instrument has read the disc ID a
screen is displayed, showing the reagent status.
5
Prepare the reagents. Deal with at least all red and orange buttons.
1. Press a reagent button to check the details.
Replace the empty reagents.
2. Press .
3. Open the main cover
4. Remove the bottle.
5. Press to confirm.
6. Remove the remaining bottles of the set.
7. Press to confirm.
8. Press .
9. Place the new bottle.
10. Press to confirm.
11. Load the remaining bottles of the set.
12. Press to proceed to the next stage in the
Prepare wizard.
Step User action
Table B-2
Steps for preparing the system (Continued)