Cedex Bio analyzer 9 General maintenance
Maintenance actions
18
Switch on the system.
The system performs internal checks and routines.
The startup phase may take a few minutes.
When the analyzer is ready for use, the Overview tab is displayed, and its status is
Standby.
19
Log on the system.
20
Choose Utilities > Maintenance > Clean Housing.
21
Press .
The maintenance definition screen is displayed.
22
Press .
A detailed maintenance definition screen is displayed.
23
Press to confirm the completion of the maintenance action.
Clean up the database
f Utilities > Maintenance > Cleanup Database
This maintenance action restores the database to the condition where the analyzer
can work at its optimum. You typically use this action when you think that the
analyzer suddenly works slow, for example when displaying information on the
screen takes long or when it takes long until an action starts after you have initiated it.
Operator time Approximately 10 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required None
a To clean up the database
1
Select the maintenance action Cleanup Database.
2
Press .
The maintenance definition screen is displayed.
3
Press .
An instructions screen is displayed while the database cleanup is performed.
4
When the maintenance action is complete, press to confirm the
completion.