EasyManuals Logo

Roche Cedex Bio User Manual

Roche Cedex Bio
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #527 background imageLoading...
Page #527 background image
Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 F-77
Cedex Bio analyzer 16 ISE troubleshooting
List of ISE flags
Calc Error
Meaning Calculation error.
o General calculation error.
o Slope or nonlinear standard curve cannot be calculated due to a calibration error.
Message ID 16
Priority 13
Possible cause o Calibrator outdated or deteriorated.
o Misplacement of the calibrator tubes.
Recommended actions o Repeat calibration with fresh calibrators.
o If the calibrators were incorrectly positioned, replace the calibrators and repeat
the calibration.
Ise Unstab
Meaning The signal from the electrode(s) was not stable during the measurement.
Message ID 18
Priority 6
Possible cause o There are air bubbles in the measuring segment.
o The concentration of sample is too low (for example, dialyzed samples).
o The ISE Reference Solution tubing is blocked or leaking.
o The pH of the sample is too low (for example, pH < 5.5).
o One or more O-rings are missing or damaged.
o There are salt deposits or there is an obstruction at the adapter of the ISE
Reference Solution bottle.
o The electrodes have not been serviced.
o The expiration date of the electrode has expired or the electrode is damaged.
o The electrodes and/or the electrode block is not dry.
o The preamplifier is defective.
Recommended actions 1. Perform the maintenance action Electrode Service and repeat the measurement.
2. Repeat the measurement with a suitable sample (where pH > 5.5 and sample
concentration within the test range) to establish that the ISE module is OK.
3. Check the position and condition of the O-rings and replace if necessary.
4. Check the adapter of the ISE Reference Solution bottle for salt deposits and
obstruction.