15 ISE maintenance Cedex Bio analyzer
ISE maintenance actions
Condition ISE tubing
f Utilities > Maintenance > Condition ISE Tubing
Conditioning the tubing with Activator ensures the proper flow of fluids.
Operator time Approximately 3 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required m Activator
a To condition the ISE tubing
1
Select the maintenance action Condition ISE Tubing.
2
Press .
The maintenance definition screen is displayed.
3
Press to start the action.
A placement list is displayed.
4
Place the Activator on the sample area position indicated on the screen.
5
Press to confirm the placement and to start the action.
A message will inform you when the maintenance action is complete.
6
Remove the Activator from the sample area.
7
Press .
Make sure that you have read and understood section Safety information on page F-46.
The following warning messages in particular are relevant:
o Injury through working solutions on page F-46.
o Infection by biohazardous materials on page F-46.