EasyManuals Logo

Roche Cedex Bio User Manual

Roche Cedex Bio
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #286 background imageLoading...
Page #286 background image
Roche Diagnostics
B-146
Operatorโ€™s Manual ยท Version 3.4
7 Configuration Cedex Bio analyzer
Introduction
Changing values
If the change of a value results in one or several other values needing to be changed, a
wizard is started when you change the first value. You recognize a wizard by the
presence of the and buttons.
a To change a value
1
Choose Utilities > Applications > Laboratory Parameters.
2
Expand the list to display the items and their values.
3
Select an item that shows a value.
4
Press .
A screen is displayed where you either can type a new value or select one by
pressing its corresponding button.
5
Type the new value or press a value button.
6
Do one of the following:
If Do this
You are in a wizard and want to change a
further value:
1. Press .
A screen is displayed for changing the
value.
2. Type the new value or press a value
button.
3. Continue with step 6.
4. When is no longer active, press
to confirm the changes.
This was the only value you want to change: 1. Press to confirm the changes.
Roche recommends not to change the calibration sequence.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio and is the answer not in the manual?

Roche Cedex Bio Specifications

General IconGeneral
BrandRoche
ModelCedex Bio
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals