Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 B-165
Cedex Bio analyzer 7 Configuration
Applications
Calculation definitions
Calculation definitions (ratio applications)
Item Values Comment
Factor Used for sample and control results.
Final result = (result * Factor) + Offset
The correlation factor is used to compensate for
differences in measuring methods of different
analyzers. By applying a factor, the final results
of the different analyzers can be directly com-
pared.
Offset The offset is used to compensate for differences
in measuring methods of different analyzers. By
adding the offset, the final results of the different
analyzers can be directly compared.
Valuation No Valuati on
Reference Range Use the limits defined as Low Limit and High
Limit.
Low Used On, Off Refers to Low Limit.
High Used On, Off Refers to High Limit.
Low Limit Results that are outside the limit are flagged.
High Limit Results that are outside the limit are flagged.
TRL Threshold value below which the results are no
longer displayed, instead it is indicated that the
value is smaller than the TRL value and the
<Test Rng flag is added.
Note that you can only change this value if TRL
Check is activated. (Utilities > Configuration >
System > TRL Check)
1...3:Flag QC Fail On, Off Choose On to ensure that all sample results are
flagged that were generated using a test whose
QC results were marked with a flag.
The initial number refers to the control that was
used for performing the QC. (See also Control
definitions.)
Table B-15 Calculation definitions
Item Values Comment
Application A
through D
Up to 5 alphanumeric
characters
Short names of the installed applications.
Coefficient w
through z
000 โ€ฆ 999 Numeric variable for use in the formula.
Formula Defined application
and coefficient letters.
Operands:
+, -, *, /, (, ), and space
The formula relates the defined applications
with the coefficients.
For example: (A * w + B) * A / B
Table B-16 Calculation definitions for ratio applications