7 Configuration Cedex Bio analyzer
Configuration
Timeout 10 ... 300 s Period of time the system will wait for a response
from the host when working in Order Query Mode.
This definition applies to routine, STAT and rerun
sample tests.
Send Result
Mode
Off The results are sent to the host manually.
Complete The results are automatically sent to the host when all
results of the order are available and accepted.
Immediate Each result is automatically sent to the host as soon as
it has been accepted.
Send Raw Data On, Off Raw data include all measurement results made
during the measuring cycles, not just the calculated
test results.
Trace On, Off The Trace function records the content of the com-
munication with the host in a trace file.
This is a troubleshooting function. Do not work with
it during normal daily operation.
The trace file can be exported as a log file (Utilities >
Export > Log Files; file format: htryyyymmd-
dhhmmss.log.)
System Device
ID
Up to 10
alphanumeric
characters.
The ID is used on reports and for communication
between the Cedex Bio analyzer and the host.
Host Device ID Up to 10
alphanumeric
characters.
The ID is used on reports and for communication
between the Cedex Bio analyzer and the host.
Item Values Comment
Table B-26
Host definitions