EasyManuals Logo

Roche Cedex Bio User Manual

Roche Cedex Bio
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #386 background imageLoading...
Page #386 background image
Roche Diagnostics
D-24
Operatorโ€™s Manual ยท Version 3.4
9 General maintenance Cedex Bio analyzer
Maintenance actions
Preparing the analyzer
13
Reinsert the sample tray.
14
Close the right service flap and the main cover.
15
Press .
The system automatically initializes the analyzer.
A message will inform you when the maintenance action is complete.
16
Press .
17
Perform an air/water calibration.
You cannot process orders unless you perform this action.
e
See Perform an air/water calibration on page D-26.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio and is the answer not in the manual?

Roche Cedex Bio Specifications

General IconGeneral
BrandRoche
ModelCedex Bio
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals