EasyManua.ls Logo

Roche Cedex Bio - Page 395

Roche Cedex Bio
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
Operator’s Manual · Version 3.4 D-33
Cedex Bio analyzer 9 General maintenance
Maintenance actions
26
Install the sample tray.
Make sure to align the guides with the pins.
Press down firmly.
27
Install and close the right service flap.
28
Move the transfer head to its rightmost position.
29
Install and close the left service flap and main cover.
Finishing the maintenance
action
30
Switch on the analyzer.
31
Wait for initialization to finish.
32
Log on to the system.
33
Insert the reagent disk.
e
See Preparing the reagent disk on page B-23

Table of Contents

Related product manuals