Roche Diagnostics
Operator’s Manual · Version 3.4 E-13
Cedex Bio analyzer 10 Messages and alarms
List of alarm messages
6021 [%s] Photometer error. The lamp voltage is outside the
range.
Restart the system. If the error persists call service.
6022 [%s] Photometer error. The internal photometer 15 V
power supply failed.
Restart the system. If the error persists call service.
6023 [%s] Photometer error. The lamp current is outside the
range.
Perform first the maintenance action <Replace Photome-
ter Lamp> and then <Abs Air/Water Calibration>. Restart
the system. If the error persists call service.
6024 [%s] Photometer error. Either the lamp fan is not running
or its speed is not correct.
Switch off the analyzer. Open the main cover and check if
the lamp fan is jamming or blocked. Restart the system. If
the error persists call service.
6025 [%s] Photometer error. Either the dark shutter motor
does not work or the shutter mechanism is jamming or
blocked.
Restart the system. If the error persists call service.
6026 [%s] Photometer error. A problem caused by the internal
circuitry has been detected.
Restart the system. If the error persists call service.
6027 [%s] Photometer error. The photometer signals are
outside the range while performing a dark measurement.
Restart the system. If the error persists call service.
6029 [%s] Photometer error. A problem caused by the internal
circuitry has been detected.
Restart the system. If the error persists call service.
6030 [%s] Initialization of the rotor failed. Switch off the analyzer. Remove the rotor and clean all
rotor measurement holes. Reinstall the rotor. Restart the
system. If the error persists call service.
6031 [%s] Photometer error. The photometer dark signal is not
stable.
Restart the system. If the error persists call service.
7001 [%d] An Instrument Control software exception error
occurred.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7002 [%d] A software error occurred. Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7004 [%d] An internal communication software error
occurred.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7006 [%s] A not specified hardware error (%d) was reported. Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7007 [%d] An internal communication error (%d) occurred. A
control unit could not be connected.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7008 [%d] A software error occurred that was caused by an
unexpected system situation.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7009 The LAN connection to the IC software failed. Check fuse F3. Check the external Ethernet cable. Restart
the system. If the error recurs export the log files with the
system remaining in this error condition. If the error per-
sists call service.
ID Message Comment
Table E-1
List of alarm messages (Continued)