Cedex Bio analyzer 10 Messages and alarms
List of alarm messages
7042 [%s] Fuse F1 blown. Switch off the analyzer. Check fuse F1. If necessary replace
fuse F1. Restart the system. If the error persists call ser-
vice.
7043 [%s] Fuse F2 blown. Switch off the analyzer. Check fuse F2. If necessary replace
fuse F2. Restart the system. If the error persists call ser-
vice.
7044 [%s] Fuse F3 blown. Switch off the analyzer. Check fuse F3. If necessary replace
fuse F3. Restart the system. If the error persists call ser-
vice.
7045 [%s] Fuse F4 blown. Switch off the analyzer. Check fuse F4. If necessary replace
fuse F4. Restart the system. If the error persists call ser-
vice.
7046 [%s] The initialization of the rotor failed. Switch off the analyzer. Check if the rotor is jamming or
blocked. Clean the rotor initialization light barrier. Restart
the system. If the error persists call service.
7051 [%s] Error rotor motor. The motor could not be started. Restart the system. If the error persists call service.
7052 [%s] Error motor X-transfer. The motor could not be
started.
Restart the system. If the error persists call service.
7053 [%s] Error motor Z-transfer. The motor could not be
started.
Restart the system. If the error persists call service.
7057 [%s] The ISE sample sensor is not calibrated or no
calibration data is available.
Perform <Initialize ISE Unit>. If the error persists call ser-
vice.
7058 [%s] ISE sample sensor could not detect fluid. Check the ISE tower and the ISE tubing for obstructions
and leaks. Ensure that the electrodes are installed cor-
rectly. If the error persists call service.
7059 [%s] The ISE sample sensor could not detect the end of
the transported fluid segment.
Check the ISE tower and the ISE tubing for obstructions
and leaks. If the error persists call service.
7060 [%s] The calibration of the ISE sample sensor failed. Check the filling level of the ISE Cal bottle. Check the ISE
tubing for obstructions and leaks. Perform <Initialize ISE
Unit>. If the error persists call service.
7061 [%s] The calibration of the ISE sample sensor failed. Check the filling level of the ISE Cal bottle. Check the ISE
tubing for obstructions and leaks. Perform <Initialize ISE
Unit>. If the error persists call service.
7062 [%s] The ISE sample sensor could not detect calibrator. Check the filling level of the ISE Cal bottle. Check the ISE
tubing for obstructions and leaks. Perform <Initialize ISE
Unit>. If the error persists call service.
7063 [%s] The ISE sample sensor could not detect the end of
the calibrator segment.
Check the ISE tubing for obstructions and leaks. Perform
<Initialize ISE Unit>. If the error persists call service.
7064 [%s] The ISE sample sensor could not detect Activator
during initialization.
Check the ISE tower and ISE tubing for obstructions and
leaks. Perform <Initialize ISE Unit>. If the error persists
call service.
7065 [%s] Calibration of the ISE reference sensor failed, no
valid calibration data are available.
Check the filling level of ISE Ref. Perform <Initialize ISE
Reference Sensor>. If the error persists call service.
7066 An internal control unit (%d) reported a wrong firmware
version.
Reinstall the system software.
7067 An internal control unit (%d) reported a checksum or an
attach error.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7068 [%d] Wash aborted due to open cover. Close the cover and restart the system.
ID Message Comment
Table E-1
List of alarm messages (Continued)