Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 E-43
Cedex Bio analyzer 12 Troubleshooting
Dealing with exceptional situations
The system does not go
to Operating status.
System status There are not enough free
cuvettes on the analyzer.
Load empty cuvette segments.
e
See Preparing cuvettes on page B-30.
Air/water calibration is
required.
Perform Abs. Air/Water Calibration maintenance
action.
e
See Perform an air/water calibration on page D-26.
In the corresponding calibrator
lot definitions of at least one of
the tests you want to perform,
the value for Last Is Water is
Off.
The Last Is Water definition defines what is used as the
zero calibrator. It is either system water (On) or a stan-
dard, usually water (Off), which would be placed in a
tube like any other calibrator.
With some tests, calibration does not work with special
zero calibrators, they need the Last Is Water definition
on.
Possible remedies:
1. Delete the current calibration order and redefine it.
2. If the Last Is Water definition is Off, make sure to
place a special zero calibrator on the analyzer when
calibrating. (This is usually water.)
If this does not help change the Last Is Water value to
On. (Calibration for this test does not work with a spe-
cial zero calibrator.)
1. Choose Wo r kp l ac e > Lot Data > Calibrators.
2. Select the entry for the test and expand the entry.
3. Select the Last Is Water entry and press .
4. Press On.
5. Press .
6. Delete the current calibration order and redefine it.
A hardware error had occurred. Some hardware conditions prevent the system from
performing tests.
1. Press
to check the alarm messages.
2. Follow the advise given in the messages.
3. Choose > Restart to restart the system.
Situation Affected
items
Possible causes Comments
Figure E-4
Troubleshooting overview (Continued)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio and is the answer not in the manual?

Roche Cedex Bio Specifications

General IconGeneral
BrandRoche
ModelCedex Bio
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals