15 ISE maintenance Cedex Bio analyzer
ISE maintenance actions
7
Clean the inside of the ISE tower with a cotton swab to remove any residual of ISE
Deproteinizer.
8
Close the left service flap and then the main cover.
9
Press to confirm the cleaning of the ISE tower.
The ISE tower is first washed.
The transfer head moves to the right side of the analyzer.
You are asked to clean the ISE tower.
10
Open the main cover and then the left service flap.
11
Clean the inside of the ISE tower with a cotton swab to remove any residual water.
12
Close the left service flap and then the main cover.
13
Press to confirm the cleaning of the ISE tower.
The ISE tower is first washed and dried and then conditioned with Activator.
A message will inform you when the maintenance action is complete.
14
Remove the Activator and ISE Deproteinizer from the sample area.
15
Press .