Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.4 F-67
Cedex Bio analyzer 15 ISE maintenance
ISE maintenance actions
14
Press to confirm completion of the action.
The system now starts to initialize the ISE Reference Solution sensor.
You are asked to insert the tubing adapter in the ISE Reference Solution bottle
and to close the ISE cover.
15
Insert the tubing adapter in the ISE Reference Solution bottle.
16
Close the ISE cover.
17
Press .
The system first pumps ISE Reference Solution through the ISE tubing. Then it
performs the Condition ISE Tubing maintenance action, and finally Initialize
ISE Unit.
A message will inform you when the maintenance action is complete.
18
Remove the Activator from the sample area.
19
Press .