Roche Diagnostics
A-42 Operator’s Manual · Version 3.4
3Hardware Cedex Bio analyzer
LEDs and color indicators
Interpreting the global action
button status indicators
Global action
button
Color Meaning
Start
Off You cannot start the measuring process.
You can start the measuring process.
Stop
Off Pressing does not have any effect.
Pressing leads to the various stop options.
Alarm
Off There are no unconfirmed alarm messages.
There is at least one unconfirmed alarm message.
You need to deal with it as soon as possible.
There is at least one unconfirmed alarm message.
You need to deal with it immediately, processing
may not be able to continue unless you do so.
An acoustic signal is sounded when an alarm is generated. You can
adjust the volume (Utilities > Configuration > System > Volum e).
Table A-5 Global action button status indicators and their meaning