Basic Instrument Functions
R&S
®
CMW500
107User Manual 1173.9463.02 ─ 06
Fig. 4-18: Info – Current State
State
Generator state, signaling generator state or measurement state, see Generator Con-
trol, Measurement Control and Control of the Cell State.
Reliability, Message, Additional Info
The reliability indicator describes the validity of measurement results and the possible
source of inaccuracies or errors.
The displayed reliability values indicate the most severe error that has been detected by
a measurement, generator or signaling application since it has been started or switched
on. A measurement returns this value also when measurement results are queried via
remote control command. When an application is stopped or switched off, the available
error information is kept. When it is re-started or switched on again, the "old" error infor-
mation is deleted.
To display all errors detected by an application, not only the most severe one, select the
application to the left and press the hotkey "Reliability List" at the bottom.
A zero in column "Reliability" indicates that no error has been detected. A detected error
is indicated via a non-zero value in column "Reliability", the corresponding text message
in column "Message" and optionally additional information like a file name or a specific
option in the last column.
For a description of all reliability indicator values, refer to chapter 5.4.4.1, "Reliability
Indicator", on page 149.
Info Dialog