Remote Control
R&S
®
CMW500
138User Manual 1173.9463.02 ─ 06
RDY: Measurement has been terminated, valid results may be available. RDY cor-
responds to the SCPI trigger state IDLE.
RUN: Measurement running (after INITiate..., READ...), synchronization
pending or adjusted, resources active or queued (see Measurement Substates). RUN
corresponds to the SCPI trigger state INITiated.
Querying the main measurement state
The main measurement state can be queried using
FETCh:<FWA>:<context>:STATe? where <FWA> denotes the firmware application
and <context> is to be replaced by the name of the measurement.
Example: FETCh:GPRF:MEASurement:POWer:STATe? (possible response: RDY).
The relationship between measurement states and measurement control commands is
shown in the following diagram:
Measurement control commands are of the following type (see also Firmware Applica-
tions):
SCPI subsys-
tem
<Application>,
e.g.
Measurement instance Context
INITiate
ABORt
STOP
:WIMax |
:GPRF
:MEASurement<i> :POWer
:EPSensor
:MEValuation
...
(depending on <Application>)
Control of the Instrument