EasyManua.ls Logo

Rohde & Schwarz CMW500 - Page 159

Rohde & Schwarz CMW500
323 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
CMW500
143User Manual 1173.9463.02 ─ 06
State Valid results? Effect of FETCh...?
OFF No FETCh...? should not be used in the OFF state as valid results are
never available.
RUN Yes/No
R&S CMW500 waits until the current measurement cycle is complete
and returns the results. In continuous repetition mode the
R&S CMW500 continues acquiring results without reaching the RDY
state.
RDY Yes/No Returns the results immediately.
Repeated FETCh...? commands
It is possible to use repeated FETCh...? commands without re-starting the measure-
ment, e.g. in order to monitor the variation in time of a measured quantity in continuous
measurement mode.
In single-shot mode where the measurement reaches the RDY state and is stopped
before the first FETCh...? is executed, the results of repeated FETCh...? com-
mands will be identical.
In continuous mode, the first FETCh...? is executed after the end of the current
measurement cycle. The following FETCh...? commands are executed as quickly
as possible, no matter whether the R&S CMW500 has acquired new results.
READ...? Command
Aborts the current measurement (if a measurement is running), starts a new measure-
ment in single-shot mode, and returns the results including the Reliability and Error Indi-
cators at the end of the measurement cycle.
READ...? does not change any of the measurement control settings. The measurement
initiated by READ...? is stopped after one single-shot (--> measurement state RDY),
however, the repetition mode itself is not changed.
READ...? can be used in all measurement states.
Control of the Instrument

Table of Contents

Related product manuals