Remote Control
R&S
®
CMW500
147User Manual 1173.9463.02 ─ 06
"Generator Control", on page 136. A signaling generator is in one of the following cell
states:
OFF: Signaling generator turned off, resources either released or (partially) reserved
from the last time the generator was turned on.
PENDing: Signaling generator turned on, but still waiting for resource allocation,
adjustment, hardware switching. No cell signal is available at the selected output
connector.
ON: Signaling generator turned on, with all necessary adjustments finished. A cell
signal to which a mobile station can synchronize is available at the selected output
connector.
The OFF and PENDing/ON states correspond to the status indication "Off" and "On" in
the generator softkeys. The relationship between generator commands and generator
states is shown in the following diagram:
Signaling generator control commands are of the following type (see also Firmware
Applications):
SCPI subsystem <Application>, e.g. Generator instance State
SOURce :GSM :SIGNaling<i> :CELL:STATe ON | OFF
Example: SOURCe:GSM:SIGNaling:CELL:STATe ON | OFF
The :CELL mnemonic is omitted in some signaling applications.