Command Reference - Base Software
R&S
®
CMW500
269User Manual 1173.9463.02 ─ 06
Return values:
<List> Comma separated list of mnemonics (and/or decimal numbers
between 0 and 65535) enclosed in brackets
Example:
STAT:OPER:TASK:A:GPRF:MEAS:POW:XEV?
Query the EVENt part of the status register for the GPRF power
measurement to check whether an event has occurred since the
last reading.
Usage: Query only
Firmware/Software: V2.0.10
STATus:
OPERation:XNTRansition <List>
STATus:OPERation:TASK:XNTRansition <List>
STATus:OPERation:TASK:A:XNTRansition <List>
STATus:OPERation:TASK:A:<netw_std>:XNTRansition <List>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:XNTRansition <List>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:XNTRansition <List>
Sets the negative transition filter. If a bit is set, a 1 to 0 transition in the corresponding bit
of the associated condition register causes a 1 to be written in the associated bit of the
corresponding event register. See also Structure of an SCPI Status Register.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to table 6-1.
Parameters:
<List> Comma separated list of mnemonics (and/or decimal numbers
between 0 and 65535) enclosed in brackets
Example:
STAT:OPER:TASK:A:GPRF:MEAS1:POW:XNTR (OFF,RDY)
Set the negative transition filter bits corresponding to the GPRF
power measurement states OFF (bit 0) and RDY (bit 3) to 1.
Firmware/Software: V2.0.10
STATus:OPERation:XPTRansition <List>
STATus:OPERation:TASK:XPTRansition <List>
STATus:OPERation:TASK:A:XPTRansition <List>
STATus:OPERation:TASK:A:<netw_std>:XPTRansition <List>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:XPTRansition <List>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:XPTRansition <List>
Sets the positive transition filter. If a bit is set, a 0 to 1 transition in the corresponding bit
of the associated condition register causes a 1 to be written in the associated bit of the
corresponding event register. See also Structure of an SCPI Status Register.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to table 6-1.
Parameters:
<List> Comma separated list of mnemonics (and/or decimal numbers
between 0 and 65535) enclosed in brackets