Getting Started
R&S
®
CMW500
13User Manual 1173.9463.02 ─ 06
Risk of injury if feet are folded out
The feet may fold in if they are not folded out completely or if the instrument is shifted.
This may cause damage or injury.
●
Fold the feet completely in or completely out to ensure stability of the instrument.
Never shift the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place anything
underneath.
●
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 500 N.
2.1.2.2 Rackmounting
The instrument can be mounted in 19" racks using an R&S ZZA-411 adapter (order num-
ber 1096.3283.00). Please note the mounting instructions supplied with the rack adapter.
Risk of instrument damage in a rack
An insufficient airflow can cause the R&S CMW500 to overheat, which may impair the
measurement results, disturb the operation, and even cause damage.
For operation in a rack, note the conditions described in chapter 2.1.2, "Positioning the
Instrument", on page 12.
Putting the Instrument into Operation