Annexes
R&S
®
CMW500
296User Manual 1173.9463.02 ─ 06
Up to 15 devices can be connected
Maximum length of the connecting cables 15 m. The length of a single connecting
cable should not exceed 2 m, if many devices are used, it should not exceed 1 m.
Wired OR if several instruments are connected in parallel
Pin assignment
Bus lines
Data bus with 8 lines D0 to D7:
The transmission is bit-parallel and byte-serial in the ASCII/ISO code. D0 is the least
significant bit, D7 the most significant bit.
Control bus with five lines:
IFC (Interface Clear): active LOW resets the interfaces of the instruments connected
to the default setting.
ATN (Attention): active LOW signals the transmission of interface messages, inactive
HIGH signals the transmission of device messages.
SRQ (Service Request): active LOW enables the connected device to send a service
request to the controller.
REN (Remote Enable): active LOW permits switchover to remote control.
EOI (End or Identify): has two functions in connection with ATN:
ATN=HIGH active LOW marks the end of data transmission.
ATN=LOW active LOW triggers a parallel poll.
Handshake bus with three lines:
DAV (Data Valid): active LOW signals a valid data byte on the data bus.
NRFD (Not Ready For Data): active LOW signals that one of the connected devices
is not ready for data transfer.
NDAC (Not Data Accepted): active LOW signals that the instrument connected is
accepting the data on the data bus.
Interfaces and Connectors