R&S PR100 User Manual
d
DISPlay
. :FSTRength?
Query whether field-strength is displayed or not
Parameters:
none
Result:
0 OFF
1 ON
Example:
DISPlay:FSTRength? -> 0
d
DISPlay
. :IFPan
. . :LEVel
. . . :RANGe <numeric_value>|MINimum|MAXimum
Sets the range of signal levels that is displayed in a panorama view. Different
levels within this range can be distinguished in the view. The top-end of the
range is equal to “DISP:IFP:LEV:REF”.
Parameters:
<numeric_value> signal-level range in dBZV
MINimum|MAXimum minimum|maximum range
Example:
DISPlay:IFPan:LEVel:RANGe 30