MEASure
. :TIME <numeric_value>|MINimum|MAXimum|DEFault
Setting of the measuring time for all measuring functions.
Remark:
The measuring time has an effect on the level detectors. When the level mode
is set to average, AVG, the measuring time determines the averaging time.
When set to peak, PEAK, this time determines the fall time. Using fast as
level method does not have any impact since it is only the current value which
is measured.
The measuring time also has an impact on the averaging time of the IF-
panorama data.
Parameters:
<numeric_value> measuring time in seconds
MINimum|MAXimum shortest/longest measuring time
DEFault use preset measuring times
Example:
MEASure:TIME 50 ms
MEASure:TIME DEF
MEASure
. :TIME? [MINimum|MAXimum]
Query of the set measuring time
Parameters:
none query of the current measuring time
MINimum|MAXimum query of the shortest/longest measuring time
Result:
Measuring time in seconds; with the default measuring time being set, DEF
will be output.
Example:
MEASure:TIME? -> 0.050000
MEASure:TIME? -> DEF