[SENSe]
. :GCONtrol
. . [:FIXed|MGC]
. . . :STEP
. . . . [:INCRement]? [MINimum|MAXimum]
Query of the MGC stepwidth.
Parameters:
none query of currently set stepwidth
MINimum|MAXimum query of smallest|largest stepwidth
Result:
MGC stepwidth in dB
Example:
GCONtrol:STEP? -> 10
[SENSe]
. :GCONtrol
. . :MODE FIXed|MGC|AUTO|AGC
Type of gain control
Parameters:
FIXed|MGC control is determined by MGC value
AUTO|AGC control is automatically generated (AGC)
Example:
GCONtrol:MODE AUTO