[SENSe]
. :MSCan
. . :DWELl <numeric_value>|MINimum|MAXimum|INFinity
Setting the dwell time. This value is mapped onto the same variable as the
frequency scan dwell variable, meaning that a change affects both scan
modes.
Parameters:
<numeric_value> dwell time in seconds
MINimum|MAXimum lowest|highest dwell time
INFinity infinite dwell time
Remark:
This command is used to set the dwell time per scan step, ie the time required
by a step, if the squelch is switched off.
Example:
SWEep:DWEL 100 ms
[SENSe]
. :MSCan
. . :DWELl? [MINimum|MAXimum]
Query of dwell time. This command is an alias of SENS:SWE:DWEL?.
Parameters:
none query of current dwell time
MINimum|MAXimum query of lowest|highest dwell time
Result:
Dwell time in seconds; 9.9E37 is output for an infinite number
Example: